Disclosed herein is an x-ray interferometer for x-ray phase contrast imaging including an x-ray source, an x-ray source grating, two x-ray phase gratings, an x-ray analyzer grating and an x-ray detector. An alternative interferometer includes a periodically structured x-ray source, two x-ray phase gratings, an x-ray analyzer grating and an x-ray detector. The phase gratings are placed much closer to the x-ray detector than to the x-ray source and the image object is positioned upstream and close to the phase gratings to achieve high sensitivity and large field-of-view simultaneously.
|
1. An x-ray interferometer for x-ray phase contrast imaging, comprising:
a) an array of equally-spaced line sources with period ps;
b) a phase grating gp1 with period p1 placed downstream of the array of line sources and at a distance l1 from the array of line sources;
c) a phase grating gp2 with period p2 place downstream of the phase grating gp1 at a distance d from the phase grating gp1;
d) an analyzer grating ga formed as an absorption grating with period pα and placed downstream of the phase grating gp2 at a distance l2 from the phase grating gp2; and
e) an x-ray detector placed downstream and close to the analyzer grating gα; wherein
f) the period of the equally-spaced line sources ps is given by:
2. The x-ray interferometer of
g) an x-ray tube source; and
h) an x-ray source grating formed as an absorption grating with the period ps placed downstream and close to the x-ray tube source.
3. The x-ray interferometer of
4. The periodically structured x-ray source according to
5. The periodically structured x-ray source according to
6. The periodically structured x-ray source according to
7. The x-ray interferometer according to
9. The x-ray interferometer according to
11. The x-ray interferometer according to
|
N/A
The present invention relates to an x-ray interferometer with an x-ray source, x-ray gratings and an x-ray detector for x-ray phase contrast imaging. More specifically, the invention relates to an x-ray interferometer that achieves high sensitivity and large field of view with less critical requirements on the small period and high aspect ratio x-ray analyzer (absorption) grating as in a Talbot-Lau interferometer of similar performance.
X-ray imaging has been used for medical diagnosis for more than a hundred years. X-ray phase contrast imaging, detecting the phase and the dark-field contrasts in addition to the absorption, has the potential to improve both the sensitivity and specificity of medical diagnosis by joint analysis of the three contrast mechanisms. Recent preclinical studies in breast imaging [Arboleda, C., et al. “Towards clinical grating-interferometry mammography.” European radiology 30.3 (2020): 1419-1425] and lung imaging [Fingerle, A. A., et al. “Imaging features in post-mortem x-ray dark-field chest radiographs and correlation with conventional x-ray and CT.” European radiology experimental 3.1 (2019): 25] demonstrate the potential clinical values of x-ray phase contrast imaging in breast imaging and lung imaging.
For daily clinical operation, the x-ray phase contrast imaging system has to be constructed with polychromatic x-ray tubes and large area medical x-ray detectors (e. g. 29 cm×24 cm for digital mammography, 43 cm×35 cm for chest x-rays) with much lower spatial resolution compared to the period of the x-ray gratings. In prior arts, the U.S. Pat. No. 7,180,979B2 describes a Talbot interferometer, utilizing the Talbot self-imaging of an x-ray phase grating to allow the use of polychromatic x-ray sources. To resolve the fine pitch of the self-imaging pattern, an absorption grating with similar period as the self-image of the phase grating is placed in front of the detector and the resulting moiré fringe is recorded by the detector, which is compatible with medical x-ray detectors with low spatial resolution. A phase stepping method is described to retrieve the absorption contrast, phase contrast and dark-field contrast images at the detector spatial resolution. For Talbot self-imaging phenomena, the source has to have sufficient spatial coherence (at least half the grating period at the phase grating), which requires synchrotron x-ray source or at least microfocus source. The U.S. Pat. No. 7,889,838B2 describes a Talbot-Lau interferometer to release the spatial coherence requirements of the x-ray source. Compared to Talbot interferometer in U.S. Pat. No. 7,180,979B2, an absorption grating is placed downstream the x-ray source. The combination of the low spatial coherence x-ray source and the source grating is equivalent to an array of line sources. Each line source provides sufficient spatial coherence for Talbot self-imaging of the phase grating. Under certain source grating period, the Talbot self-imaging of the phase grating from each line source is aligned in phase and added up (in intensity) constructively, although the line sources are spatially incoherent with each other. The combination of the low coherence x-ray source and the source grating can be replaced by a micro-structured x-ray source, providing an array of equally spaced sources, with a period satisfying Talbot-Lau condition as described in U.S. Pat. No. 7,889,838B2 and U.S. Pat. No. 9,719,947B2.
The maximum interferometer sensitivity of a Talbot-Lau interferometer is proportional to the distance between the phase grating and the analyzer grating and inversely proportional to the period of the analyzer grating. The interferometer sensitivity at the image object is a fraction of the maximum sensitivity. If the image object is placed upstream the phase grating, the fractional factor is the ratio of the distance between the source grating and the image object to the distance between the source grating and the phase grating. If the image object is placed downstream the phase grating, the fractional factor is the ratio of the distance between the image object and the analyzer grating to the distance between the phase grating and the analyzer grating. For a given x-ray imaging application, assuming sufficiently large gratings to cover the entire x-ray detector, the field of view is a fraction of the detector area, where the fractional factor is the square of the ratio of the distance between the x-ray source and the image object to the source-to-detector distance. For clinical diagnoses, such as digital mammography and chest x-ray radiography, to achieve large field of view, the patient position has to be close to the detector. Under the patient to detector distance and source-to-detector distance restrictions, the period of the analyzer grating has to be sufficiently small to get high interferometer sensitivity. Since a thick layer of high atomic number material (e. g. 50 μm Au for 30 keV x-rays and 200 μm Au for 60 keV x-rays) is required to sufficiently absorb hard x-ray photons, high aspect ratio x-ray gratings are necessary in Talbot-Lau interferometer to achieve compact, highly sensitive x-ray phase contrast imaging system. Fabrication of the small period and high aspect ratio analyzer gratings is a major challenge in the development of Talbot-Lau interferometers for clinical applications.
Analyzer grating free x-ray interferometers have been developed, including an x-ray polychromatic far field interferometer with 3 phase gratings [Miao, H., et al. “A universal moiré effect and application in X-ray phase-contrast imaging.” Nature physics 12.9 (2016): 830-834] and dual phase grating x-ray interferometers [Kagias, M., et al. “Dual phase grating interferometer for tunable dark-field sensitivity.” Applied Physics Letters 110.1 (2017): 014105. Yan, A., Wu, X., and Liu, H. “Quantitative theory of x-ray interferometers based on dual phase grating: fringe period and visibility.” Optics Express 26.18 (2018): 23142-23155. Ge, Y., et al. “Dual phase grating based X-ray differential phase contrast imaging with source grating: theory and validation.” Optics Express 28.7 (2020): 9786-9801.]. Although, the sensitivity of the 3-grating polychromatic far field interferometer is the highest among all reported grating-based interferometers, the large distance required between the detector and the grating next to the detector and the large inter-grating distances makes the field of view small. Moreover, the intrinsic fringe period of such an interferometer is typically not resolvable by medical x-ray detectors. The small field of view together with the incompatibility with large area medical x-ray detectors limits its clinical potential, particularly for full field imaging such as in applications of breast imaging and chest radiography. The dual-phase grating x-ray interferometers suffer from small fringe period that cannot be resolved by large area medical detectors, small focal spot requirement of the x-ray source, and the low interferometer sensitivity. These drawbacks make it difficult to develop a practical imaging system for clinical diagnosis.
This disclosure provides an x-ray phase contrast imaging system using an analyzer grating of much larger period and lower aspect ratio than that used in a Talbot-Lau interferometer of similar performance. The system is compatible with polychromatic, low spatial coherence medical x-ray tubes and large area medical x-ray detectors for large field of view imaging, which is required for applications, such as in lung imaging, breast imaging and abdomen imaging.
The system consists of
The combination of a standard polychromatic medical x-ray source and the source grating can be replaced by a periodically structured x-ray source. An alternative system consists of
The two x-ray phase gratings form a universal moiré pattern at the plane of the analyzer grating. The period of the analyzer grating is selected to convert the universal moiré pattern to lower spatial frequency fringe via intensity moiré effect. The resulted intensity moiré pattern can be resolvable by the x-ray detector, which requires a period of at least 3 times the detector spatial resolution, preferably ≥5 times the detector resolution.
The spatial coherence requirement of the interferometer limits the spot size of the x-ray source to a few μm or a few tens of μm. X-ray microfocus source cannot meet the requirement of small focal spot size and high output power at the same time for fast data acquisition. This is solved in this disclosure by using either a source grating (absorption grating) placed downstream and close to the x-ray source or a periodically structured x-ray source. The source grating converts the x-ray source to a group of line sources, where the coherence of each line source meets the coherence requirement. To achieve high fringe visibility, the source period is selected such that the universal moiré pattern formed by each individual source adds up constructively.
The image object is designed to be placed upstream and close to the two phase gratings. The interferometer sensitivity is proportional to the distance between the source grating (or structured source) and the image object, and inversely proportional to the source period (ps). The two phase gratings are placed close to the analyzer grating and the detector to simultaneously increase the interferometer sensitivity and the field of view. A certain distance between the phase gratings and the analyzer grating is required to achieve high fringe visibility. When the two phase gratings are selected to have the same period, the source and the analyzer gratings have the same period. Alternatively, the two phase gratings can be selected to have different periods to reduce the source period and increase the analyzer grating period. Decreased source grating period improves the interferometer sensitivity. In some applications, the analyzer grating period can be increased large enough to allow the use of large area x-ray grid made of lead to simplify the instrumentation and reduce the cost.
The embodiments of the present invention will become better understood with reference to the following drawings. It is noted that, for purpose of illustrative clarity, certain elements in various drawings may not be drawn to scale. These drawings depict exemplary embodiments of the disclosure, but should not be considered to limit its scope. Preferred examples and embodiments are described hereinafter with reference to the accompanying drawings, wherein:
Various embodiments of the disclosure are discussed in details below. While specific implementations are discussed, it should be understood that this is done for illustration purposes only. A person skilled in the relevant art will recognize that other components and configurations may be used without parting from the spirit and scope of the disclosure. Thus, the following description, drawings and examples are illustrative and are not to be construed as limiting.
In a first embodiment of the present invention, the x-ray interferometer is constituted by a medical x-ray source 100, an x-ray absorption grating called a source grating 200, a first x-ray phase grating 211, a second x-ray phase grating 212, an x-ray absorption grating called an analyzer grating 220, and an x-ray detector 300 (refer to
In a second embodiment, the conventional medical x-ray source and the source grating is replaced by a structures x-ray source 100-SS (Refer to
The x-ray source 100 (Refer to
The x-ray phase gratings 211 and 212 (Refer to
The analyzer grating 220 is an x-ray absorption grating, consisting of a low x-ray absorption substrate 221, a group of low absorption grating teeth 222 and a group of high absorption grating teeth 223 (Refer to
The period of the source grating 200 (Refer to
Where L, L2 and D are the distance from the source grating 200 or the structured source 100-SS to the analyzer grating 220, the distance from the phase grating 212 to the analyzer grating 220, and the distance between the two phase gratings 211 and 212, respectively, as illustrated in
The period of the analyzer grating 220 (Refer to
Where L1 is the distance from the source grating 200 or the structured source 100-SS to the phase grating 211, as illustrated in
The image object 400 is designed to be upstream and close to the phase grating 211. The interferometer sensitivity at the object is η=LO/ps, where LO is the distance from the source grating 200 or the structured source 100-SS to the image object 400, as illustrated in
For a specific application, given the source-to-detector distance restriction, the period and position of the phase gratings 211 and 212 (Refer to
The embodiments of the invention are best described by application examples as described below.
The interferometer (Refer to
Variations of the grating parameters from the designed values are well tolerated. A 2% fringe visibility degradation allows at least, for the source grating 200 or the analyzer grating 220, +5% variation of the duty cycle; or for the phase gratings 211 or 212, +10% variation of the grating height, or +5% variation of the duty cycle, or +20% variation of the phase rising (or falling) width.
The grating parameters can be easily controlled within the acceptable range during the grating fabrication process. In one embodiment of the grating fabrication, the source grating 200 of 30.1 μm period is fabricated by deep reactive ion etching (DRIE) of Si to 300 μm in depth. The Si grating teeth width is controlled to around 7.5 μm by controlling the grating teeth width during the lithography process and the DRIE process. The trenches are then filled with Au via electrodeposition. The phase gratings 211 and 212 of 900 nm period are patterned by i-line stepper and deep etched by DRIE of Si. By controlling the DRIE process, the grating 211 height is controlled to 9±0.9 μm, the average Si teeth width is controlled to 585±45 nm, and the phase rising (or falling) width is controlled to 225±45 nm. The grating 212 height is controlled to 12.5±1.25 μm, the average Si teeth width is controlled to 585±45 nm, and the phase rising (or falling) width is controlled to 225±45 nm. Au is deposited via electroplating to fill the trenches to complete the fabrication of phase gratings 211 and 212. The analyzer grating 220 of 30.1 μm period is fabricated by DRIE of Si to 300 μm in depth. The Si grating teeth width is controlled to around 15 μm by controlling the grating teeth width during the lithography process and the DRIE process. The trenches are then filled with Au via electrodeposition.
Large area x-ray absorption gratings made of Au are costly. In this example, the analyzer grating period is enlarged by using two phase gratings with different period, so that a lead x-ray grid can be used as the analyzer grating. The interferometer (Refer to
A 2% fringe visibility degradation allows at least, for the source grating 200 or the analyzer grating 220, ±5% variation of the duty cycle; or for the phase gratings 211 or 212, ±10% variation of the grating height, or ±5% variation of the duty cycle, or ±20% variation of the phase rising (or falling) width. The grating parameters can easily be controlled to the tolerant range during the fabrication process.
The interferometer (Refer to
A 2% fringe visibility degradation allows at least, for the source grating 200 or the analyzer grating 220, ±5% variation of the duty cycle; or for the phase gratings 211 or 212, ±10% variation of the grating height, or ±5% variation of the duty cycle, or ±20% variation of the phase rising (or falling) width. The grating parameters can easily be controlled to the tolerant range during the fabrication process.
Patent | Priority | Assignee | Title |
Patent | Priority | Assignee | Title |
7180979, | Dec 26 2002 | X-ray imaging system and imaging method | |
7889838, | Jun 06 2005 | PAUL SCHERRER INSTITUT | Interferometer for quantitative phase contrast imaging and tomography with an incoherent polychromatic x-ray source |
9348067, | Jun 16 2009 | Koninklijke Philips Electronics N V | Tilted gratings and method for production of tilted gratings |
9532760, | Apr 24 2012 | Siemens Healthcare GmbH | X-ray device |
9719947, | Oct 31 2013 | SIGRAY, INC | X-ray interferometric imaging system |
9763634, | May 22 2013 | Siemens Healthcare GmbH | Phase-contrast X-ray imaging device |
9881710, | Mar 27 2009 | Koninklijke Philips Electronics N V | Achromatic phase-contrast imaging |
20050286680, | |||
20090092227, | |||
20120020461, | |||
20120057677, | |||
20150055744, | |||
20150243397, | |||
20160066873, | |||
EP1447046, | |||
EP3139156, | |||
RU2584247, | |||
RU2666153, | |||
WO2010150136, | |||
WO2013160153, |
Executed on | Assignor | Assignee | Conveyance | Frame | Reel | Doc |
Date | Maintenance Fee Events |
Oct 06 2021 | BIG: Entity status set to Undiscounted (note the period is included in the code). |
Oct 18 2021 | MICR: Entity status set to Micro. |
Date | Maintenance Schedule |
Nov 14 2026 | 4 years fee payment window open |
May 14 2027 | 6 months grace period start (w surcharge) |
Nov 14 2027 | patent expiry (for year 4) |
Nov 14 2029 | 2 years to revive unintentionally abandoned end. (for year 4) |
Nov 14 2030 | 8 years fee payment window open |
May 14 2031 | 6 months grace period start (w surcharge) |
Nov 14 2031 | patent expiry (for year 8) |
Nov 14 2033 | 2 years to revive unintentionally abandoned end. (for year 8) |
Nov 14 2034 | 12 years fee payment window open |
May 14 2035 | 6 months grace period start (w surcharge) |
Nov 14 2035 | patent expiry (for year 12) |
Nov 14 2037 | 2 years to revive unintentionally abandoned end. (for year 12) |